Methods
8-27
ADVIA
2120/2120i PEROX SHEATH encases the sample stream as the two
fluids pass through the flowcell.
The light-scatter and the light-absorption signals are detected and electronically
amplified. After processing, the following information is available:
PEROX Rate
histogram displays the arrival rate of cells in the Perox channel.
PEROX X
histogram displays the absorption values for all cells. Information
from the noise area is excluded.
PEROX Y
histogram displays the light scattering values for all events, including
those from the noise area.
Noise-Lymph
histogram provides the light scattering values for the noise,
lymphocyte, and large unstained cells (LUC) areas. Information from the
monocyte, neutrophil, and eosinophil areas is excluded.
PEROX Cytogram
is formed from the paired light scatter and absorption data.
Cluster analysis identifies the individual populations.
Perox Rate Histogram
The PEROX Rate
histogram shows the
uniformity of the cell-
counting rate.
The rate histogram data consists of 50 points, one taken every 200 milliseconds.
Each point represents the number of valid cells counted during the last 200
milliseconds.
Perox X Histogram
PEROX X histogram
displays the absorption
values for all cells.
Information from the noise
area is excluded.
Increasing channel numbers on the PEROX X histogram correspond to greater
intensity of peroxidase staining.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...