Regulatory Information
9-41
Parameter
Expected
Value
MN (cells/
μ
L)
0-10
PMN (cells/
μ
L)
0-4
Performance Characteristics: Precision
Within run precision on the ADVIA 120 CSF Assay was evaluated by
performing replicate aspirations of 16 CSF patient samples with adequate
volume. Volume permitting, CSF samples were aspirated 2–10 times. Results
from the 16 samples were pooled giving a total sample size of n=65.
Precision of CSF WBC Counts (n=65)
Parameter Mean
SD
CV
WBC 44.6 5.0 11.2
Precision of CSF RBC Counts (n=65)
Parameter Mean
SD
CV
RBC 12.9
4.1 31.7
Precision of CSF WBC Differential (n=65)
Parameter Mean SD
CV
% Neut
12.5
3.2
25.5
% Lymph
74.9
5.1
6.9
% Mono
8.3
3.2
39.3
% MN
83.2
3.6
4.3
% PMN
16.8
3.6
21.2
# Neut
4.8
1.1
23.6
# Lymph
33.8
4.0
11.7
# Mono
4.1
1.2
30.1
# MN
37.9
4.7
12.4
# PMN
6.6
2.0
29.9