Troubleshooting the Analyzer
5-3
Adjusting the Length of the Sample Probe
After each aspiration, the wash block (
1
) is lowered and the probe is washed
within the block. For a proper wash, the terminal end of the probe (
2
) must be
between the upper (
3
) and lower (
4
) ports of the wash block when the block is at
its lowest position.
CORRECT PROBE ALIGNMENT
PROBE MUST BE REPLACED
PROBE ALIGNMENT REQUIRED
1.
Look at the terminal end of the probe in the lowered wash block. If the probe
is too short, it must be replaced.
2.
If the probe is too long, remove the probe and, using a sharp scalpel or a
single-edged razor blade, cleanly cut the probe to the appropriate length. Be
very careful to make the cut straight without burrs.
Do not cut the probe on
an angle.
3.
Replace the probe and check the length.
Adjusting the Pneumatic Regulators
1.
To gain access to the knobs, snap open the white panel on the right side of
the analyzer. Using a Phillips-head screwdriver, remove the small green
panel.
2.
At the Utilities menu, select
Analyzer Status
.
3.
Monitor the pneumatic values on the screen as you adjust the appropriate
knob.
20 IN.-HG VACUUM
REGULATOR KNOB
40 PSI PRESSURE
REGULATOR KNOB
5 PSI PRESSURE
REGULATOR KNOB
20 PSI PRESSURE
REGULATOR KNOB
Turn knobs clockwise to increase values and counterclockwise to decrease
values.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...