Troubleshooting the Analyzer
5-41
Flowcell release knob (
1
)
Front of analyzer
6.
Hold the flowcell by the red threaded fitting located at the top of the flowcell,
then gently lift the flowcell out of the optics assembly.
CAUTION
To avoid getting fingerprints on the glass windows of the flowcell, always
hold the flowcell by the metal slides.
To install the flowcell
1.
Hold the flowcell by its red threaded fitting and place it gently into the optics
assembly.
2.
Tighten the release knob to lock the flowcell into position.
3.
Reconnect the hydraulic lines to the CFM. Reconnect the flowcell tube
assembly fitting to the appropriate hydraulic valve, V24. Hand tighten the
connection enough to prevent leaks.
IMPORTANT
To avoid incorrect patient data, make sure that each hydraulic line is attached to
the correct CFM nipple.
To check analyzer performance
1.
Bring the analyzer to the Ready to Run mode, then run a saline primer.
2.
Check the flowcell connections for leaks and verify saline background on the
Startup tab.
3.
At the Procedures menu, select the
Align
Optics
tab.
4.
Select the Info button for help.
5.
At the Logs menu, select
Cal/Gain Logs
and check gain factors. Adjust gains
if necessary.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...