8. Loosen the 2 screws where the slots are on the Autoslide cart bracket so you
can move the support structure.
Figure 5. Autoslide cart bracket with loosened screws
9. Move the support structure forward up to 0.3 m (1 ft), ensuring that the
electrical cables are not stretched to far.
1
Blood sample line
2
Pressure line
3
Ground strap
4
Vacuum line
Figure 6. Moved Support Structure
NOTE:
The ground strap and interface cable remain connected to the Autoslide.
Disconnecting the sample blood line, pressure line, and vacuum lines allows you
to move the Autoslide approximately 0.3 m (1 ft).
10.
To reconnect the Autoslide, move the Autoslide module back into position.
11.
Tighten the clamp screws.