Flags
6-17
Platelet Clumps (PLT-CLM, NW)
Definition
Presence of clumped platelets is suspected.
The Platelet Clumps flag is triggered if the Clumps Count in the PLT Clumps
region of the Perox cytogram is greater than 150. The PLT count is flagged to
alert the user that the platelet count may not be accurate due to the presence of
platelet clumps.
The PLT Clump area of the Perox cytogram originates in the Noise area and
extends upward to the right of the lymphocyte population. The number of events
obtained from this area is called the Clumps Count.
The Clumps Count is excluded from the WBCP and differential results.
There is no severity level associated with this flag.
Perox Cytogram
Normal Sample
Perox Cytogram
PLT-CLM flag
1 = PLT Clumps area
Results Flagged
WBCP, PLT, PCT, MPV, and PDW
Possible Causes
The following list of sample-related causes may not contain all conditions that
could cause this flag.
Multiple occurrences of this flag, especially for consecutive samples, can indicate
a system problem. However, isolated instances of this flag are usually sample
related.