7-96
Status Line messages
Check the length of the sample ID.
Unknown Host Test Number xx
A workorder from the host contains a test that is not defined in the Test
Dictionary or contains an incorrect host number.
Corrective Action
Modify the Test Dictionary or correct the host test table. Do not modify the Test
Dictionary while samples remain in the database.
V
Vacuum Out of Range
System vacuum is out of range.
Possible Cause
1
Vacuum gauge
is out of range
Corrective Action
Check the system vacuum gauge and adjust the regulator
if necessary.
Setpoint 20" Hg
±
1.
2
Vacuum leak
a
Check for crimped or disconnected hydraulic or
pneumatic lines. Straighten or reconnect lines as
necessary.
b
Check the waste container to make sure that the spigot
is closed and the cap is on securely.
Vacuum Out of Range - Stop
System vacuum is out of range.
Possible Cause
1
Vacuum gauge
is out of range
Corrective Action
Check the system vacuum gauge and adjust the regulator
if necessary.
Setpoint 20" Hg
±
1.
2
Vacuum leak
a
Check for crimped or disconnected hydraulic or
pneumatic lines. Straighten or reconnect lines as
necessary.
b
Check the waste container to make sure that the spigot
is closed and the cap is on securely.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...