Methods
8-41
RBC Volume histogram (microcytic sample)
1
Microcytic region
2
Normocytic region
3
Macrocytic region
4
60 fL marker
5
120 fL marker
In samples with increased numbers of microcytic red blood cells, the histogram
curve shifts to the left, indicating an increase in the percentage of the cells with
volumes less than 60 fL.
RBC Volume histogram (macrocytic sample)
1
Microcytic region
2
Normocytic region
3
Macrocytic region
4
60 fL marker
5
2120/2120i fL marker
In samples with increased numbers of macrocytic red blood cells, the histogram
curve shifts to the right, indicating an increase in the percentage of the cells with
volumes greater than 2120/2120i fL.
RBC Volume histogram (anisocytosis)
1
Microcytic region
2
Normocytic region
3
Macrocytic region
4
60 fL marker
5
120 fL marker
RDW is monitored as an indication of anisocytosis, and the results are flagged if
the RDW exceeds 16%. Note that two specimens with the same RDW value can
have different degrees of microcytosis and macrocytosis.