
Status Line messages
7-33
Corrective Action
a
Remove the rack from the shuttle position at the front of the output queue.
b
Select
Start
at the analyzer touchpad to restart autosampler operation.
Autosampler - Output Shuttle Failed
There is no rack in the output shuttle position at the end of shuttle motion. The
probable cause is the removal of the rack by the user before the shuttle stops.
Autosampler - Output Shuttle Not Empty
The autosampler has attempted to move a rack to the output shuttle, but the
shuttle is not empty.
Corrective Action
Remove rack from the output shuttle.
Autosampler - Pushpin Extend Denied
To prevent the collision of autosampler components, the system did not execute
the command to extend the autosampler pushpin.
Autosampler - Pushpin Motion Denied, Reset Autosampler
The autosampler attempted an inappropriate pushpin motion. An autosampler
reset is required.
Corrective Action
1
Select
Off
at the analyzer touchpad.
2
Wait about 1 minute and select
On
to restart the analyzer.
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Pushpin Retract Denied
To prevent the collision of autosampler components, the system did not execute
the command to retract the autosampler pushpin.
Autosampler - Pushpin Failed to Extend, Reset Autosampler
The autosampler pushpin failed to extend properly.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...