Welcome to the ADVIA 2120/2120i Hematology System
1-15
Stand-alone waste container
Automatic Waste Removal
How the ADVIA 2120/2120i Hematology System Works
Blood samples can be aspirated through the:
•
Autosampler
•
Manual closed-tube sampler
•
Manual open-tube sampler
After a sample (
4
) is aspirated, it is drawn into the shear valve (
5
). As the shear
valve rotates, it “shears” or divides the sample into aliquots for the different types
of tests.
24
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18 19 20
21
22
23
BASO
RETIC
RBC
HGB
PEROX
5
4
V74
V2
V72
V73
V47
V1
Rinse Trapped in Lines
Lines that are Dry
Lines with Sample
Line with Vacuum
Conductivity Detector
Rinse Source
Probe
Vacuum Source
The reagents and sample segments are delivered to their respective reaction
chambers for mixing and aspiration. (PEROX 2 and PEROX 3 reagents are
delivered directly to the perox reaction chamber.)