Status Line messages
7-9
Possible Cause
1
Power failure
at the analyzer.
Corrective Action
Select
Off
on the analyzer touchpad and verify that the
power plug is connected and secure. Reconnect if
necessary and select
On
at the analyzer touchpad.
2
Faulty or
disconnected
Ethernet cable
(from the
analyzer - J2
Workstation
connection to
the computer).
Verify that the Ethernet cable connection is secure.
Replace if necessary.
3
The computer
networking
configuration
has been
corrupted.
Follow these steps to exit the ADVIA 2120/2120i
software, turn off the analyzer, and then restart:
a
Select
Shut Down ADVIA
in the Log On/Off
window.
b
Select
Off
at the analyzer touchpad.
c
Select
CTRL+ALT+DELETE
and log back onto the
system to restart the software.
d
Restart the analyzer by selecting
On
at the analyzer
touchpad.
e
If the problem persists, call Siemens Service for
assistance.
Anisocytosis - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
Anisocytosis - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
Aspiration Failed - Probe Clog - Alarm
Conductivity detectors did not detect a sample within the 20 seconds of
aspiration. The number of consecutive occurrences of this condition has met the
criterion set for an alarm message.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...