Status Line messages
7-29
2
Wait about 1 minute and select
On
to restart the analyzer.
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Needle Extend Denied
To prevent the collision of autosampler components, the system did not execute
the command to extend the autosampler needle.
Autosampler - Needle Failed to Extend, Reset Autosampler
The autosampler needle failed to extend properly.
Possible Cause
1
Critical
Sampler
Error.
Corrective Action
a
Open the autosampler access door.
b
Close the autosampler access door. The autosampler
will reset.
c
If the problem persists, select
Off
on the analyzer
touchpad.
d
Wait about 1 minute and select
On
to restart the
analyzer.
e
If the problem persists, call Siemens Service for
assistance.
2
40 PSI line is
out of range.
a
Check pressure and adjust if necessary.
b
Check for pinched air lines supplying needle cylinder.
c
On the analyzer touchpad, select
Off
.
d
Wait about 1 minute and select
On
to restart the
analyzer.
e
If the problem persists, call Siemens Service for
assistance.
3
Needle is
bent.
Replace needle.
4
Needle sensor
failed.
Call Siemens Service for assistance.
Autosampler - Needle Failed to Retract, Reset Autosampler
The autosampler needle failed to retract properly.
Possible Cause
1
Critical
Sampler
Corrective Action
a
Open the autosampler access door.
b
Close the autosampler access door. The autosampler
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...