8-44
Methods
1
Low-angle light scatter (2° to 3°)
2
High-angle light scatter (5° to 15°)
3
Mie map containing RBCs
4
Platelets detected in RBC method
Using the Mie theory of light scattering for homogeneous spheres, the low-angle
and high-angle light scatter signals for each cell are transformed into volume and
hemoglobin concentration values.
The RBC map shows the relationship between the light-scatter measurements and
the cell-by-cell characteristics of volume and hemoglobin concentration. The
map grid encompasses RBC volumes between 30 fL and 180 fL and hemoglobin
concentrations between 19 g/dL and 49 g/dL.
For a normal specimen, the RBC population appears in the center of the map.
However, since this map is nonlinear, visually evaluating red cell morphology
can be difficult. Use the V/HC cytogram, which provides a linear presentation of
cell volume and hemoglobin concentration.
RBC Volume / Hemoglobin (V / HC) Cytogram
The Volume/Hemoglobin Concentration (V/HC) cytogram is a linear version of
the RBC map that appears on the RBC cytogram. On the V/HC cytogram,
hemoglobin concentration is plotted along the x axis and cell volume is plotted
along the y axis. Only red blood cells appear on this cytogram.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...