Troubleshooting the Analyzer
5-21
Shuttle line from UFC assembly (
3
)
Sample line from syringe (
2
)
Sheath line from hydraulic valve (
1
)
V15 - perox or V19 - RBC
Cleaning the Flowcells off the Analyzer: Step 7
Checking Analyzer Performance
IMPORTANT
An inspection scope must be available for this procedure. Do not remove the
flowcells unless you are trained in how to align and focus them.
1.
Bring the analyzer to the
Ready to Run
mode.
2.
At the Operations menu, select
Startup
,
select
Refresh
, and verify the
background count.
3.
Check the flowcell connections for leaks.
4.
Check the focus and alignment of the flowcell.
5.
Check the gains.
6.
Run a whole-blood primer, then run controls to verify analyzer performance.
7.
If the control results are acceptable to the laboratory, no additional action is
required and normal operation can be resumed.
Backflushing the Perox Flowcell and Shuttle and Reaction
Chambers
Materials required:
•
Flowcell cleaning kit, PN 113-B711-01
•
EZ KLEENBeaker
•
Deionized water
Analyzer mode: Ready to Run
BIOHAZARD
All products or objects that come in contact with human or animal body fluids
should be handled, before and after cleaning, as if capable of transmitting
infectious diseases. Wear facial protection, gloves, and protective clothing.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...