Status Line messages
7-89
Sample xxx Is Not Transmitted to the Host
There is a transmission problem to the host computer. The message specifies the
Sid# of the nontransmitted samples. Another message will follow identifying the
specific problem.
Scan Denied - Reset Autosampler
Barcode reader is unable to scan.
Possible Cause
1
Faulty barcode
reader cabling
Corrective Action
a
At the analyzer touchpad, select
Off
.
b
Wait about 1 minute and select
On
to restart the
analyzer.
c
If the problem persists, call Siemens Service for
assistance.
2
Faulty barcode
reader
See
1a
,
b
,
c
SD Not Defined For Test# xx Control# xxyyzzz
No standard deviation is defined for this test and for this control in the Control
Dictionary.
Corrective Action
Define the SD in the Control Dictionary.
Selected Printer Not Available - Result Report for SID#:
The printer is not responding to a print request.
Corrective Action
Check that the printer is powered on and has paper. Refer to the Message Log for
details.
Selector Valve Error
Potential software communication error or a hardware failure associated with the
selector-valve assembly.
Possible Cause
1
Hardware
failure
Corrective Action
a
Reset the analyzer by selecting
Utilities, Exerciser,
Indicators, Analyzer Reset
.
b
After about 1 minute, if the status line does not display
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...