ADVIA Autoslide Slide Maker Stainer
10-33
7.
Draining (AN7).
8.
Dispensing rinsing solution (distilled water) (DN7).
9.
Draining (AN4).
10.
Dispensing methanol (DN4).
11.
Slide removal.
Needle Position
Action
DN1 (Undil. Stain 1)
32
1
DN2 (Buffer)
22
0
DN3 (Dil. Stain 2)
8
1
DN4 (Methanol)
31
0
DN7 (Rinse)
24
2
AN2 22
0
AN3 8
1
AN4 31
1
AN7 24
2
NOTE:
Needles AN2 and DN2 in position 22 are unused in this protocol.
Methanol Volume (mL) 4
Stain Volume (mL)
3.6
Dilution Ratio (%)
25
Duration Long
Principle of the Procedure
The stained smear is examined microscopically; the nucleus and the cytoplasm of
neutrophils, lymphocytes, monocytes, eosinophils, and basophils show a
characteristic blue or red coloration. Cells are then manually differentiated and
quantified.
Cell Type
Cell Component
Color
Neutrophils
Nuclei
Vibrant deep purple
Cytoplasm
Pink
background
Granules
Light
purple-violet
Lymphocytes
Nuclei
Distinct deep purple
Cytoplasm
Sky blue to deep blue
Monocytes
Nuclei
Light purple with distinct spaces
between chromatin strands
Cytoplasm
Dull blue-gray with lightly stained
Granules
Fine pink to violet and evenly
distributed
Eosinophils Nuclei
Pale
purplish-blue and well defined
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...