Status Line messages
iv
Restart the analyzer by selecting
On
at the
analyzer touchpad.
2
Faulty power
supply, fuse,
cable, Valve
Driver 2 Board,
or Can Bus
Scrambler
Board
Call Siemens Service for assistance.
W
Waste Container - Full
Waste level in the container has reached the level sensor indicating that the
container is full. When this error message is displayed, the aspirate plate will not
function, nor will the Ready light illuminate.
Possible Cause
1
The waste
container is
full.
Corrective Action
Follow the steps in Emptying the Waste Container.
2
Level sensor
cable is
disconnected.
Verify connection is secure.
3
Faulty level
sensor.
Call Siemens Service for assistance.
WBC Substitution - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
WBC Substitution - Stop
The number of consecutive occurrences of this condition has
autosampler halts at the same time the stop message occurs. The specified stop
criterion is user definable in the Customize Menu, System Setup, Alarm/Stop
Criteria. The counter associated with the Stop Criteria is reset when the
Start/Stop button is selected.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...