ADVIA Autoslide Slide Maker Stainer
10-49
Event Text
Severity
Probable Cause
Suggested Action
ASL - Impossible to rinse
Failure
Unable to rinse sample or drop
deposit lines; Other command is
processing.
Perform a mechanical initialization; If
problem persists, call Siemens.
ASL - Internal software
error
Failure
Software initialization problem or
incorrect data entered.
Enter correct information; Run a
mechanical initialization cycle. Call your
Siemens Representative if not
resolved.
ASL - Internal
synchronization error
Failure
Autoslide internal synchronization
error.
Perform a mechanical initialization. If
problem persists, call Siemens.
ASL - Needle plate error
Failure
Staining needle positioning
problem.
Open the covers and identify the
problem. Close covers. Refer to Tips 2
& 3. If needles hitting slides, call
Siemens.
ASL - Operation
completed in error
Failure
Run Stopped - Autoslide
command completed with an error.
Check Message Log - Detail screen for
cause, correct and perform Mechanical
Initialization.
ASL - Printer error
Failure
Printer ribbon not advancing or
jammed.
Open front panel, inspect printer ribbon
- clear jam or replace ribbon as
needed. Ensure that there are no glass
particles in the printer area. Run a
Mechanical Initialization cycle. Cycle
autoslide power if necessary.
ASL - Rack Error - Stop
Failure
General rack problem, could be
low on racks.
Replenish racks in dispenser, check for
jammed rack or full ejection tray. Run a
reset smearer. (See Tips above)
ASL - Rack Handling
error
Failure Hardware
failure in rack handler
track.
Check for jammed racks or debris. Run
a reset smearer. (See Tips above)
ASL - Slide Carrier error
Failure
Slide did not make it to the carrier
in time, or loaded incorrectly.
Clear slide and inspect area for glass
particles or debris then refer to
Troubleshooting Tip # 2.
ASL - Smearing Error -
Stop
Failure
Smearer door was open. Run
stopped if Alarm/Stop criteria set.
Close smearer door, reset smearer
(see Tip #2).
ASL - Smearing module
door is open
Failure
Smearing module door is open.
Close the smearing module door, reset
smearer if necessary.
ASL - Smearing module
not primed
Failure
Smearing module is not primed.
Run prime smearer hydraulic cycle and
reset smearer. Perform Daily Startup if
necessary.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...