Troubleshooting the Analyzer
5-29
Replacing a Perox Check Valve
Time: Installation and Checkout - 15 minutes
Materials required: Check valve, PN 556-1190-01
Analyzer mode: Standby
BIOHAZARD
All products or objects that come in contact with human or animal body fluids
should be handled, before and after cleaning, as if capable of transmitting
infectious diseases. Wear facial protection, gloves, and protective clothing.
The operator should follow the recommendations to prevent the transmission of
infectious agents in health-care settings as recommended for potentially
infectious specimens in
Protection of Laboratory Workers from Infectious
Disease Transmitted by Blood, Body Fluids, and Tissue
, 2nd edition; Approved
Guideline (1997) Document M29-A, National Committee for Clinical Laboratory
Standards (NCCLS). This document contains complete information on user
protection and it can be used as reference material for instructions on laboratory
safety.
1.
Remove the old check valve from the Perox line.
2.
Install the check valve into the line
with the direction of flow toward
the reagent pump and away from
the reagent bottle, as shown by the
red arrow in the figure to the right.
Make sure not to damage the tube
sleeve on the valve.
3.
Prime the reagent lines. Visually verify that the lines are priming.
4.
Run controls.
Replacing the Front Shear Face or the Shear Valve Rotor
Materials required:
•
ADVIA TESTpoint controls
•
Calibrators (as required)
•
Front shear valve face, PN 067-0558-01
•
Shear-valve rotor, PN 067-0876-01
•
Whole blood
Time: Replacement - 5 minutes
Checkout - 15 minutes
Analyzer mode: Standby
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...