
9-44
Regulatory Information
CSF RBC Low Level Linearity
Pool
(% of reference level)
Observed
RBC
Expected
RBC
Absolute
Deviation
%
Deviation
0.00% 0
0
0
0.05% 2
1
1
0.1% 4
3
1
0.2% 9
6
3
1.0% 34
29
5
10.0% 308
288
20
7
100% (reference level)
2880
2880
0
0
NOTE:
Although RBC linearity has been demonstrated up to 2880 RBC/
μ
L,
samples with greater than 1500 RBC/
μ
L should be diluted to ensure the accuracy
of the white count.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...