6-38
Flags
System Related
Sample Related
1.
Check perox reagents.
2.
Check peroxidase hydraulics and sample
delivery.
3.
Check delivery of SHEATH/RINSE.
4.
Check PEROX SHEATH delivery.
5.
Check the pressure and vacuum readings.
6.
Check the perox reaction chamber
temperature.
7.
Check perox flowcell alignment.
8.
Check perox gains.
•
Neonatal samples
•
Aged blood sample
•
Malaria parasites
•
Lyse-resistant RBCs
Perox Power Low (PX-PL, PX)
Definition
The Perox Power Low flag is triggered if the tungsten-halogen light intensity in
the Perox channel is less than 90.
Results Flagged
WBCP, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EOS,
#EOS, %LUC, and #LUC
Corrective Action
1.
Perform the sheath reagent check.
2.
Check perox lamp alignment.
3.
Replace the tungsten-halogen lamp.
Perox Saturation (PX-SAT, XS)
Definition
The Perox Saturation flag is triggered if the events counted in the Saturation area
are more than 10% of Perox signals
The Saturation area of the Perox cytogram is located in channels 97 through 99
on the x axis.