The measurement starts immediately.
Parameters:
<State>
ON
Continuous measurements.
OFF
Single measurements.
Example:
INIT2:CONT OFF
Turns on a single measurement.
Manual operation:
See
"Continuous Bargraph / Single Bargraph"
See
See
See
INITiate<n>[:IMMediate]
The command initiates a new measurement.
In case of a single measurement, the R&S
ESRP stops measuring when it has
reached the end frequency. When you start a continuous measurement, it stops only if
you abort it deliberately.
If you are using trace modes MAXHold, MINHold and AVERage, previous results are
reset when you restart the measurement.
In single sweep mode, you can synchronize to the end of the measurement with
*OPC
,
*OPC?
or
*WAI
. In continuous sweep mode, synchronization to the end of the mea-
surement is not possible. Thus, it is not recommended that you use continuous sweep
mode in remote control, as results like trace data or markers are only valid after a sin-
gle sweep end synchronization.
Example:
INIT2:CONT OFF
Selects single measurement mode.
SWE:COUN 20
Forms an average over 20 measurements.
INIT2;*WAI
Starts the measurement and waits for the end of the complete
measurement.
Usage:
Event
Manual operation:
See
See
11.4.1.4
Final Measurements and Automated Test Sequence Control
.............................................................................................611
[SENSe:]DETector<t>:FMEasurement
.............................................................................. 611
.......................................................................................612
........................................................................................ 612
Remote Commands in Receiver Mode