SYSTem:REBoot
This command reboots the instrument, including the operating system.
Usage:
Event
SYSTem:SHUTdown
This command shuts down the instrument.
Usage:
Event
SYSTem:SPEaker:VOLume
<Volume>
This command defines the volume of the built-in loudspeaker for demodulated signals.
Parameters:
<Volume>
Range:
0 to 1
*RST:
0.5
Example:
SYST:SPE:VOL 0
Switches the loudspeaker to mute.
SYSTem:TIME
<Time>
This command sets the internal clock. The sequence of entry is hour, minute, second.
Only user accounts with administrator rights are able to set the time.
Parameters:
<Time>
0 to 23, 0 to 59, 0 to 59
Example:
SYST:TIME 12,30,30
Usage:
SCPI confirmed
Manual operation:
See
SYSTem:TIME:TSTamp?
This command queries the time and date from the internal real time clock.
Return values:
<TimeStamp>
Example:
SYST:TIME:TSTamp?
Usage:
Query only
SYSTem:VERSion?
This command queries the number of the SCPI version, which is relevant for the instru-
ment.
System Configuration