Remote Control
R&S
®
ESRP
759
User Manual 1175.7445.02 ─ 08
The third-order intercept is calculated from the level spacing between the normal mark-
ers and the delta markers.
Suffix:
<n>
.
Selects the measurement window.
<m>
irrelevant
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:MARK:FUNC:TOI ON
Switches on the measurement of the third-order intercept.
Manual operation:
See
See
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult?
This command queries the results for the third order intercept point measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m>
irrelevant
Return values:
<TOI>
Third order intercept point.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:TOI ON
Switches the intercept measurement.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:TOI:RES?
Outputs the measured value.
Usage:
Query only
Manual operation:
See
See
11.5.1.11
Measuring Statistics
The following commands configure and perform statistics measurements.
Remote Commands in Spectrum Analyzer Mode