Remote Control
R&S
®
ESRP
617
User Manual 1175.7445.02 ─ 08
[SENSe:]DEMod:SQUelch:LEVel
<Threshold>
This command defines the threshold for selective demodulation. Squelching must be
active for this command to work (see
All signals below the threshold are not demodulated.
This command in only available if the R&S
FSV-B3 option (Audio Demodulation) is
installed.
Parameters:
<Threshold>
The threshold level as a percentage of the display height.
Range:
0 to 100
*RST:
60
Example:
DEM:SQU:LEV 80
Sets the squelch level to 80% of the displayed signal.
Usage:
SCPI confirmed
Manual operation:
See
See
[SENSe:]DEMod:SQUelch[:STATe]
<State>
This command turns selective demodulation at the marker position on and off.
A video trigger is automatically activated with the same level as the squelch level, any
other trigger or gate settings are deactivated.
This command in only available if the R&S
FSV-B3 option (Audio Demodulation) is
installed.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
DEM:SQU ON
Signals below the level threshold are not sent to the audio out-
put.
Usage:
SCPI confirmed
Manual operation:
See
See
11.4.1.7
Spectrogram Control
.................................................................................... 618
...............................................................................................627
Remote Commands in Receiver Mode