System Configuration
R&S
®
ESRP
487
User Manual 1175.7445.02 ─ 08
Language
Comment
8564E
8565E
8566A
Command sets A and B are available. Command sets A and B differ in the rules
regarding the command structure.
8566B
8568A
Command sets A and B are available. Command sets A and B differ in the rules
regarding the command structure.
8568A_DC
Uses DC input coupling by default if supported by the instrument
8568B
Command sets A and B are available. Command sets A and B differ in the rules
regarding the command structure.
8568B_DC
Uses DC input coupling by default if supported by the instrument
8591E
Compatible to 8594E
8594E
Command sets A and B are available. Command sets A and B differ in the rules
regarding the command structure.
For details on the GPIB languages, see
Chapter 11.18, "GPIB Commands of HP Mod-
els 856xE, 8566A/B, 8568A/B and 8594E"
Remote command:
IF Gain (Norm/Puls) ← Compatibility Mode ← GPIB ← General Setup
Configures the internal IF gain settings in HP emulation mode due to the application
needs. This setting is only taken into account for resolution bandwidth < 300
kHz.
NORM
Optimized for high dynamic range, overload limit is close to reference level.
PULS
Optimized for pulsed signals, overload limit up to 10
dB above reference level.
This softkey is only available if an HP language is selected via the "GPIB Language"
softkey (see
Remote command:
Sweep Repeat (On/Off) ← Compatibility Mode ← GPIB ← General Setup
Controls a repeated sweep of the E1 and MKPK HI/HL HP model commands (for
details on the commands refer to
sweep is "OFF", the marker is set without sweeping before.
Note:
In single sweep mode, switch off this softkey before you set the marker via the
E1 and MKPK HI/HL commands to avoid sweeping again.
This softkey is only available if an HP language is selected via the "GPIB Language"
softkey (see
Remote command:
Instrument Setup and Interface Configuration – SETUP Key