Spectrum Measurements
R&S
®
ESRP
286
User Manual 1175.7445.02 ─ 08
Figure 6-8: CCDF of white noise
A red line indicates the ideal Gaussian distribution for the measured amplitude range
(white noise).
The displayed amplitude range is indicated as "Mean Pwr + <x dB>"
In addition to the histogram, a result table is displayed containing the following informa-
tion:
●
Number of samples used for calculation
●
For each displayed trace:
Mean
Mean power
Peak
Peak power
Crest
Crest factor (peak power – mean power)
0,01 %
Level values over 0,01 % above mean power
0,1 %
Level values over 0,1 % above mean power
Measurements