Remote Control
R&S
®
ESRP
702
User Manual 1175.7445.02 ─ 08
Manual operation:
See
Defining Weighting Filters
The following commands define weighting filters for ACLR measurements.
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel
........................................................ 702
[SENSe:]POWer:ACHannel:FILTer:ALPHa[:ALL]
............................................................... 702
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ALTernate<channel>
......................................... 702
[SENSe:]POWer:ACHannel:FILTer:ALPHa:CHANnel<channel>
.......................................... 703
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ACHannel
....................................................... 703
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALTernate<channel>
........................................ 703
[SENSe:]POWer:ACHannel:FILTer[:STATe]:CHANnel<channel>
......................................... 703
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALL
................................................................ 704
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel
<Alpha>
This command defines the roll-off factor for the weighting filter of the adjacent channel.
Parameters:
<Alpha>
<numeric value>
*RST:
0,22
Example:
POW:ACH:FILT:ALPH:ACH 0,35
Manual operation:
See
[SENSe:]POWer:ACHannel:FILTer:ALPHa[:ALL]
<Alpha>
This command defines the roll-off factor for the weighting filter of all channels.
Parameters:
<Alpha>
<numeric value>
*RST:
0,22
Example:
POW:ACH:FILT:ALPH:ALL 0,35
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ALTernate<channel>
<Alpha>
This command defines the roll-off factor for the weighting filter of the specified alternate
channel.
Suffix:
<channel>
.
1...11
the alternate channel
Parameters:
<Alpha>
<numeric value>
*RST:
0,22
Example:
POW:ACH:FILT:ALPH:ALT3 0,35
Sets the alpha value for the weighting filter for the alternate
channel 3 to 0,35.
Remote Commands in Spectrum Analyzer Mode