Remote Control
R&S
®
ESRP
768
User Manual 1175.7445.02 ─ 08
The number of results required for the calculation of average is defined with
.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:AVER ON
Switches on the calculation of average.
AVER:COUN 200
Sets the measurement counter to 200.
INIT;*WAI
Starts a sweep and waits for the end.
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MODE
<Mode>
This command selects absolute or relative power measurement in zero span.
The reference power for relative measurement is defined with
MARKer<m>:FUNCtion:SUMMary:REFerence:AUTO ONCE
. If the reference power
is not defined, the value 0 dBm is used.
Suffix:
<n>
.
Selects the measurement window.
<m>
irrelevant
Parameters:
<Mode>
ABSolute | RELative
*RST:
ABSolute
Example:
CALC:MARK:FUNC:SUMM:MODE REL
Switches the power measurement in zero span to relative.
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd
<State>
This command turns the peak hold function for the active power measurement in zero
span on and off.
If on, the measurement results show only the maximum power that has been mea-
sured for each sweep point.
The peak hold function is reset by switching it off and on again.
Remote Commands in Spectrum Analyzer Mode