Receiver Mode
R&S
®
ESRP
139
User Manual 1175.7445.02 ─ 08
Remote command:
Bargraph Detector
Opens a submenu to select the detector for the bargraph result display.
Each detector you select adds another bargraph to the result display. Up to four bar-
graphs at the same time are possible.
For more information see
Remote command:
[SENSe:]DETector:RECeiver[:FUNCtion]
Couple to Scan Trace ← Bargraph Detector
Couples or decouples the bargraph detector and scan detectors.
If on, the R&S
ESRP does as follows.
●
Turns on a scan detector for every active bargraph.
If you add a new bargraph detector, the corresponding scan trace is automatically
turned on.
●
Matches the scan trace number to the number of the bargraph.
●
Matches the color of the scan trace to a particular bargraph detector.
If you couple bargraph and scan trace, the R&S
ESRP replaces the detectors of all
other active scan traces with the new detector type.
Remote command:
DISPlay:BARGraph:TCOupling[:STATe]
Measurement Time
Defines the measurement time for a scan and the bargraph.
Remote command:
Demod
Opens a submenu to configure AM or FM demodulation.
Demod (On Off) ← Demod
Turns demodulation at the receiver frequency on and off.
Remote command:
AM / FM ← Demod
Selects AM or FM demodulation.
Remote command:
Squelch ← Demod
Defines the minimum level for the signal to be demodulated.
For more information see
Chapter 5.2.4, "AF Demodulation"