User Manual 1175.7445.02 ─ 08
3.2.1.9
Performing a Self Alignment and a Self Test
Operating temperature
Before performing this functional test, make sure that the instrument has reached its
operating temperature (for details, refer to the data sheet).
Performing a self alignment
1. Press the [SETUP] key.
2. Press the "Alignment" softkey.
3. Press the "Self Alignment" softkey.
Once the system correction values have been calculated successfully, a message
is displayed.
To display the alignment results again later
●
Press the [SETUP] key.
●
Press the "Alignment" softkey.
●
Press the "Show Align Results" softkey.
Performing a self test
The self test does not need to be repeated every time the instrument is switched on. It
is only necessary when instrument malfunction is suspected.
1. Press the [SETUP] key.
2. Press the "More" softkey.
3. Press the "Service" softkey.
4. Press the "Selftest" softkey.
Once the instrument modules have been checked successfully, a message is dis-
played.
3.2.1.10
Checking the Supplied Options
The instrument may be equipped with both hardware and firmware options. In order to
check whether the installed options correspond to the options indicated on the delivery
note, proceed as follows.
1. Press the [SETUP] key.
2. Press the "System Info" softkey.
3. Press the "Ve Options" softkey.
A list with hardware and firmware information is displayed.