Remote Control
R&S
®
ESRP
725
User Manual 1175.7445.02 ─ 08
[SENSe:]LIST:RANGe<range>:SWEep:TIME
This command sets the sweep time for the specified range.
Suffix:
<range>
.
1...20
range
Parameters:
<SweepTime>
Allowed values depend on the ratio of span to RBW and RBW to
VBW. For details refer to the data sheet.
*RST:
0.27 s
Example:
LIST:RANG1:SWE:TIME 1
Sets the sweep time for range 1 to 1 s.
Manual operation:
See
[SENSe:]LIST:RANGe<range>:SWEep:TIME:AUTO
This command activates or deactivates the automatic sweep time setting for the speci-
fied range.
Suffix:
<range>
.
1...20
range
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
LIST:RANG3:SWE:TIME:AUTO OFF
Deactivates the sweep time auto mode for range 3.
Manual operation:
See
[SENSe:]LIST:RANGe<range>:TRANsducer
<TransducerName>
This command sets a transducer for the specified range. You can only choose a trans-
ducer that fulfills the following conditions:
●
The transducer overlaps or equals the span of the range.
●
The x-axis is linear.
●
The unit is dB.
Suffix:
<range>
.
1...20
range
Parameters:
<TransducerName> 'string' = name of the transducer
Example:
SENS:LIST:RANG1:TRAN 'test'
Sets the transducer called "test" for range 1.
Remote Commands in Spectrum Analyzer Mode