Remote Control
R&S
®
ESRP
688
User Manual 1175.7445.02 ─ 08
Measuring the Third Order Intercept Point
............................................................758
..............................................................................................759
............................................................................. 767
.................................................................................. 779
.................................................................................. 783
.............................................................................................. 788
11.5.1.1
Controlling Power Measurements
The following commands control power measurements (channel power, occupied
bandwidth etc.) in general.
CALCulate<n>:MARKer<m>:FUNCtion:POWer:SELect
..................................................... 688
CALCulate<n>:MARKer<m>:FUNCtion:POWer:RESult?
.................................................... 689
CALCulate<n>:MARKer<m>:FUNCtion:POWer[:STATe]
.....................................................692
CALCulate<n>:MARKer<m>:FUNCtion:POWer:MODE
...................................................... 692
.................................................................................................693
................................................................................ 693
[SENSe:]POWer:ACHannel:PRESet:RLEVel
.....................................................................694
CALCulate<n>:MARKer<m>:FUNCtion:POWer:SELect
<MeasType>
This command selects – and switches on – the specified power measurement type in
the window specified by the suffix <n>.
The channel spacings and channel bandwidths are configured in the
SENSe:POWer
subsystem.
Note
: If CPOWer is selected, the number of adjacent channels (
) is set to 0. If ACPower is selected, the number of adjacent
channels is set to 1, unless adjacent-channel power measurement is switched on
already.
The channel/adjacent-channel power measurement is performed for the trace selected
with
The occupied bandwidth measurement is performed for the trace on which marker 1 is
positioned. To select another trace for the measurement, marker 1 is to be positioned
on the desired trace by means of
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<MeasType>
ACPower | CPOWer | MCACpower | OBANdwidth | OBWidth |
CN | CNO
ACPower
Adjacent-channel power measurement with a single carrier sig-
nal
Remote Commands in Spectrum Analyzer Mode