Remote Control
R&S
®
ESRP
961
User Manual 1175.7445.02 ─ 08
Example:
SYSTem:HPC HP
Usage:
SCPI confirmed
Manual operation:
See
SYSTem:IFGain:MODE
<Mode>
Configures the internal IF gain settings in HP emulation mode due to the application
needs. This setting is only taken into account for resolution bandwidth < 300 kHz and is
only available if a HP language is selected using
.
Parameters:
<Mode>
NORM | PULS
NORM
Optimized for high dynamic range, overload limit is close to ref-
erence level.
PULS
Optimized for pulsed signals, overload limit up to 10 dB above
reference level.
*RST:
NORM
Example:
SYST:IFG:MODE PULS
Usage:
SCPI confirmed
Manual operation:
See
SYSTem:TWALking
<Mode>
Defines how remote control programs are evaluated, in particular multiple commands
in one command line.
Note that the default setting is
NORM
; however, the setting defined by this command is
retained after a [preset] (*RST) operation.
Parameters:
<Mode>
NORM
SCPI-conform behaviour
The first command is always referenced to the root node. Sub-
sequent commands, however, are referenced to the same tree
level as the previous command in a message unit.
ENHanced
If subsequent commands are not found on the same tree level
as the previous command, the SCPI parser searches "back-
wards up the tree" of defined SCPI commands. If the command
is found at a higher node, that command is used. Only if the
command is not found anywhere, an error message is returned.
*RST:
setting is retained
Network Connection