Quick Start
R&S
®
ESRP
98
User Manual 1175.7445.02 ─ 08
Figure 3-11: Measurement of two equally-leveled RF sinusoidal signals with the resolution band-
width which corresponds to the frequency spacing of the signals
Matching generator and R&S
ESRP frequencies
The level drop is located exactly in the center of the screen only if the generator
frequencies match the frequency display of the R&S
ESRP exactly. To achieve
exact matching, the frequencies of the generators and the R&S
ESRP must be
synchronized.
5. Set the resolution bandwidth to 100 kHz. To do so, in the bandwidth menu, press
the "Res BW Manual" softkey and enter
100 kHz
.
It is no longer possible to clearly distinguish the two generator signals.
Basic Measurement Examples