result$ = Space$(100)
CALL InstrRead(analyzer, result$, 100, retCount)
Debug.Print "Marker Count Freq: ";result$
END SUB
REM *************************************************************************
Working with a Fixed Reference Point
The following example is based on a signal at 100 MHz with a level of -20 dBm. Thus,
the harmonics of the signal are located at 200 MHz, 300 MHz, etc. For high-quality sig-
nal sources, these harmonics may be located outside the dynamic range of the
R&S
ESRP. Nevertheless, to measure the harmonic suppression, the level setting
must be changed to higher sensitivity when measuring the harmonics. In this case, it
may be necessary to suppress the carrier by using a notch filter in order to prevent the
RF input of the R&S
ESRP from being overloaded.
Thus, two measurements with different level settings are performed in the following
example. First, a high reference level is used on the carrier frequency, and then a low
reference level is used on the frequency of the third harmonic.
The default setting of the R&S
ESRP for measurements ("SetupInstrument") is also
used as a starting point here, after which adaptations for the measurement are carried
out.
REM *************************************************************************
Public Sub RefFixed()
Dim retCount as Long
CALL SetupInstrument 'Default setting
'--------- Measuring the reference point ----------------------------------
CALL InstrWrite(analyzer,"INIT:CONT OFF")
'Switch to single sweep
CALL InstrWrite(analyzer,"CALC:MARK:PEXC 6DB")
'Define peak excursion
CALL InstrWrite(analyzer,"CALC:MARK:STAT ON")
'Activate marker 1
CALL InstrWrite(analyzer,"CALC:MARK:TRAC 1")
'Set marker 1 to trace 1
CALL InstrWrite(analyzer,"INIT;*WAI") 'Perform sweep with sync
CALL InstrWrite(analyzer,"CALC:MARK:MAX")
'Set marker 1 to 100 MHz
CALL InstrWrite(analyzer,"CALC:DELT:FUNC:FIX ON")
'Define reference point
'----- Setting frequency, level and bandwidth for measuring harmonics ------
CALL InstrWrite(analyzer,"FREQ:CENT 400MHz;Span 1MHz")
'Set freq of 3rd harmonic
CALL InstrWrite(analyzer,"BAND:RES 1kHz")
'Set suitable RBW
CALL InstrWrite(analyzer,"SWEEP:TIME:AUTO ON")
'Couple sweep time
CALL InstrWrite(analyzer,"INP:ATT:AUTO ON")
'Select more sensitive level setting
CALL InstrWrite(analyzer,"DISP:WIND:TRAC:Y:RLEV -50dBm")
Controlling the R&S