Remote Control
R&S
®
ESRP
898
User Manual 1175.7445.02 ─ 08
INSTrument:COUPle:MARKer
<State>
This command turns synchronization of the marker frequency between measurement
channels on and off.
Parameters:
<State>
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
NONE
Example:
INST:COUP:MARK ALL
Synchronizes the receiver frequency and the marker frequency.
INSTrument:COUPle:PRESel
<State>
This command turns synchronization of the preselector state between measurement
channels on and off.
Note that it is not possible to turn off the preselector in receiver mode. Thus, the prese-
lector is always turned on when you enter receiver mode.
Parameters:
<State>
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
ALL
Example:
INST:COUP:PRES ALL
Synchronizes the preselector configuration.
INSTrument:COUPle:PROT
<State>
This command turns synchronization of the input protection between measurement
channels on and off.
(
➙
Parameters:
<State>
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
NONE
Example:
INST:COUP:PROT ALL
Synchronizes the "10 dB Min" parameter.
System Configuration