Receiver Mode
R&S
®
ESRP
159
User Manual 1175.7445.02 ─ 08
Unit
Relative unit
dBµV
dBµV/MHz
dBµV/m
dBµV/mMHz
(Available for active transducers only.)
dBmV
dBmV/MHz
dBµA
dBµA/MHz
dBµA/m
dBµA/mMHz
(Available for active transducers only.)
dBpW
dBpW/MHz
dBpT
dBpT/MHz
The conversion to 1
MHz bandwidth is realized via the pulse bandwidth of the selected
resolution bandwidth.
Example:
Conversion example for dbµV:
MHz
1
[MHz]
B
log
20
P[dBµV]
]
P[dBµV/MHz
imp
P
= Displayed level
B
imp
= Pulse bandwidth of the selected RBW
If you are using another unit, replace "dBµV" with the corresponding unit.
The conversion is also possible when a transducer defines the used unit.
Remote command:
Grid Range / Grid Min Level
Defines the scale of the vertical diagram axis.
The display ranges go from 10 to 200 dB in 10-dB steps. Invalid entries are rounded off
to the nearest valid value.
"Grid Range"
Defines the level display range for the scan diagram.
"Grid Min
Level"
Defines the minimum level of the display range.
Remote command:
Defining the range of the grid:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
Defining the minimum level displayed on the axis:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:BOTTom
Common Measurement Settings