Remote Control
R&S
®
ESRP
690
User Manual 1175.7445.02 ─ 08
ACPower
Adjacent-channel power measurement
Results are output in the following sequence, separated by com-
mas:
Power of transmission channel
Power of lower adjacent channel
Power of upper adjacent channel
Power of lower alternate channel 1
Power of upper alternate channel 1
Power of lower alternate channel 2
Power of upper alternate channel 2
The number of measured values returned depends on the num-
ber of adjacent/alternate channels selected with
With logarithmic scaling (RANGE "LOG"), the power is output in
the currently selected level unit; with linear scaling (RANGE "LIN
dB" or "LIN %"), the power is output in W. If
is set to "REL", the adjacent/alternate-channel
power is output in dB.
AOBW (AllOccupiedBandWidth)
Returns the occupied bandwidth, as well as the position and
level of the temporary markers T1 and T2 used to calculate the
occupied bandwidth.
The syntax of the result is:
<Occupied bandwidth>,<T1 x-value>,<T1 y-value>,<T2 x-
value>,<T2 y-value>
CN
Measurement of carrier-to-noise ratio
The carrier-to-noise ratio in dB is returned.
CN0
Measurement of carrier-to-noise ratio referenced to 1 Hz band-
width.
The carrier-to-noise ratio in dB/Hz is returned.
CPOWer
Channel power measurement
In a Spectrum Emission Mask measurement, the query returns
the power result for the reference range, if this power reference
type is selected.
With logarithmic scaling (RANGE LOG), the channel power is
output in the currently selected level unit; with linear scaling
(RANGE LIN dB or LIN %), the channel power is output in W.
PPOWer
Power of the highest peak
In a Spectrum Emission Mask measurement, the query returns
the power result for the reference range, if this power reference
type is selected.
Remote Commands in Spectrum Analyzer Mode