Receiver Mode
R&S
®
ESRP
150
User Manual 1175.7445.02 ─ 08
14
N
N
2
15
L
L
6
18
3
12
8
13
+5 V
+5 V
1
ESH3-Z5
Pin
Filter
PE fl
DGND
PE flo
DGND
Adapter Pin
(9-pol
25-pol)
2
6
3
8
1
R&S Receiver
Pin
Figure 5-6: Connection from R&S
ESRP to R&S ESH3-Z5
14
N
N
14
15
L
L
15
18
18
12
12
13
+5 V
+5 V
13
R&S Receiver
Pin
ENV216
Pin
Filter
PE fl
DGND
PE flo
DGND
Adapter Pin
(9-pol
25-pol)
2
6
3
8
1
Figure 5-7: Connection from R&S
ESRP to R&S ESH216
14
N
N
14
15
L1
L1
15
16
16
17
17
12
GND
GND
12
13
+5 V
+5 V
13
L2
L3
L2
L3
Filter
ENV4200 / ENV432
Pin
2
6
7
4
8
1
Adapter Pin
(9-pol
25-pol)
R&S Receiver
Pin
Figure 5-8: Connection from R&S
ESRP to R&S ENV4200 or R&S ENV432
To control the phase selection and PE simulating network of the V-Networks
R&S
ESH2-Z5, R&S
ESH3-Z5, R&S
ENV4200 and R&S
ENV432, the +5
V supply volt-
age and some control lines have to be routed through the wall of the shielded room.
You can also use a direct connection without a filter, e.g. when you use the R&S
ESRP
in a shielded room. In that case, you can use the following cables.
●
R&S ESH2-Z5: EZ-5, EZ-13, EZ14
●
R&S ESH3-Z5: EZ-6, EZ-14
●
R&S ENV216: EZ14
●
R&S ENV4200 and R&S ENV432: EZ-14, EZ-21
5.2.6
Transducers
A transducer is often connected ahead of the R&S
ESRP both during the measure-
ment of useful signals and EMI and converts the useful or interference variable such as
field strength, current or RFI voltage into a voltage across 50
Ω
. Because most trans-
ducers such as antennas, probes or current probes have a characteristic frequency
response, it is necessary to correct the measurement results by the frequency charac-
teristics of the transducer. These characteristics are defined in a transducer factor or
transducer sets. The transducer factor can be stored in the R&S
ESRP and automati-
cally has the correct unit during level measurement.
Measurement Basics