CALCulate<n>:LIMit<k>:ACPower:ACHannel[:RELative]:STATe
<State>
This command activates the limit check for the relative limit value of the adjacent chan-
nel when adjacent-channel power measurement is performed. Before this command,
the limit check must be activated using
CALCulate<n>:LIMit<k>:ACPower[:
.
The result can be queried with
CALCulate<n>:LIMit<k>:ACPower:ACHannel:
. Note that a complete measurement must be performed between switching on
the limit check and the result query, since otherwise no correct results are available.
Suffix:
<n>
.
Selects the measurement window.
<k>
irrelevant
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:LIM:ACP:ACH 30DB, 30DB
Sets the relative limit value for the power in the lower and upper
adjacent channel to 30 dB below the channel power.
CALC:LIM:ACP:ACH:ABS -35DBM, -35DBM
Sets the absolute limit value for the power in the lower and
upper adjacent channel to -35 dBm.
CALC:LIM:ACP ON
Switches on globally the limit check for the channel/adjacent
channel measurement.
CALC:LIM:ACP:ACH:STAT ON
Switches on the check of the relative limit values for adjacent
channels.
CALC:LIM:ACP:ACH:ABS:STAT ON
Switches on the check of absolute limit values for the adjacent
channels.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:LIM:ACP:ACH:RES?
Queries the limit check result in the adjacent channels.
CALCulate<n>:LIMit<k>:ACPower:ACHannel:RESult
This command queries the result of the limit check for the upper/lower adjacent chan-
nel when adjacent channel power measurement is performed.
If the power measurement of the adjacent channel is switched off, the command pro-
duces a query error.
Suffix:
<n>
.
Selects the measurement window.
<k>
irrelevant
Remote Commands in Spectrum Analyzer Mode