Remote Control
R&S
®
ESRP
625
User Manual 1175.7445.02 ─ 08
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:NEXT
This command positions the specified marker on the next peak level of the vertical
axis. The search includes frames above and below the current marker position. It does
not change the horizontal position of the marker.
The command is available for the spectrogram.
Suffix:
<n>
.
1...4
Selects the measurement window.
<m>
1...16
Selects the marker.
Example:
CALC:MARK2:SGR:Y:MAX
Activates and positions marker 2 on the peak level of the y-axis.
CALC:MARK2:SGR:Y:MAX:NEXT
Positions marker 2 on the next higher level.
Usage:
Event
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum[:PEAK]
This command positions the specified marker on the current peak level of the vertical
axis. The search is performed over all frames. It does not change the horizontal posi-
tion of the marker.
The command is available for the spectrogram.
Suffix:
<n>
.
1...4
Selects the measurement window.
<m>
1...16
Selects the marker.
Example:
CALC:MARK2:SGR:Y:MAX
Activates and positions marker 2 on the peak level of the y-axis.
Usage:
Event
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum:ABOVe
This command positions the specified marker on the next minimum level of the vertical
axis. The search includes only frames above the current marker position. It does not
change the horizontal position of the marker.
The command is available for the spectrogram.
Suffix:
<n>
.
1...4
Selects the measurement window.
Remote Commands in Receiver Mode