Remote Control
R&S
®
ESRP
833
User Manual 1175.7445.02 ─ 08
*RST:
POINts
Example:
CALC:MARK:X:SSIZ STAN
Sets the measured value step size.
Manual operation:
See
See
CALCulate<n>:MARKer<m>:Y:PERCent
<Probability>
This command positions the selected marker to the given probability.
If you use the command in combination with a delta marker, that delta marker is turned
into a normal marker.
Note
: The command is only available for CCDF measurements. You can query the
associated level value with
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<Probability>
Range:
0 to 100
Default unit: PCT
Example:
CALC1:MARK:Y:PERC 95PCT
Positions marker 1 to a probability of 95 %.
Manual operation:
See
CALCulate<n>:THReshold
<Threshold>
This command defines a threshold value for the marker peak search.
A threshold line is automatically turned on.
Suffix:
<n>
.
irrelevant
Parameters:
<Threshold>
The unit depends on
*RST:
(STATe to OFF)
Example:
CALC:THR -82DBM
Sets the threshold value to -82 dBm.
Manual operation:
See
CALCulate<n>:THReshold:STATe
<State>
This command turns the threshold line for the marker peak search on and off.
Suffix:
<n>
.
irrelevant
Remote Commands in Spectrum Analyzer Mode