Remote Control
R&S
®
ESRP
609
User Manual 1175.7445.02 ─ 08
ABORt
This command stops a measurement and resets the trigger system.
Example:
ABOR;INIT:IMM
Aborts the measurement and restarts it.
Usage:
SCPI confirmed
Manual operation:
See
See
See
HOLD
This command interrupts a running scan measurement.
To resume the scan, use
Example:
HOLD
Interrupts the scan.
Manual operation:
See
See
See
INITiate<n>:CONMeas
This command resumes a scan that was interrupted by a transducer stop at the current
reciever frequency.
If the scan was interrupted by the
command, you have to resume it with
Example:
INIT2:CONT OFF
Selects single scan mode.
SWE:COUN 20
Forms an average over 20 scans.
INIT2;*WAI
Starts the measurement and waits for the end of the 20 scans.
INIT2:CONM;*WAI
Continues the measurement (next 20 sequences) and waits for
the end.
Usage:
Event
Manual operation:
See
INITiate<n>:CONTinuous
<State>
This command turns single measurements on and off.
Remote Commands in Receiver Mode