Remote Control
R&S
®
ESRP
747
User Manual 1175.7445.02 ─ 08
Remote commands exclusive to EMI measurements:
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:STATe
........................................... 747
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:STATe
<State>
This command turns the EMI measurement marker functionality on and off.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<State>
ON | OFF
Performing a Peak Search
Commands useful to perform EMI measurements described elsewhere:
●
CALCulate<n>:DELTamarker<m>:AOFF
●
CALCulate<n>:DELTamarker<m>:TRACe
●
●
Remote commands exclusive to EMI measurements:
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:DETector
................................747
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DETector
....................................... 748
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DWELl
.......................................... 748
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:PSEarch:AUTO
..............................749
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:DETector
<Detector>
This command selects the detector for a specific deltamarker during the final evalua-
tion.
If the deltamarker hasn't been active yet, the command also turns the deltamarker on.
Suffix:
<n>
.
irrelevant
<m>
deltamarker
Parameters:
<Detector>
OFF
turns the deltamarker off
AVER
average detector
CAV
CISPR Average detector
CRMS
RMS Average detector
Remote Commands in Spectrum Analyzer Mode