Remote Control
R&S
®
ESRP
693
User Manual 1175.7445.02 ─ 08
[SENSe:]POWer:TRACe
<TraceNumber>
This command assigns the channel/adjacent channel power measurement to the indi-
cated trace. The corresponding trace must be active, i.e. its state must be different
from blank.
Note:
The measurement of the occupied bandwidth (OBW) is performed on the trace
on which marker 1 is positioned. To evaluate another trace, marker 1 must be posi-
tioned to another trace with
.
Parameters:
<TraceNumber>
1 to 6
Example:
POW:TRAC 2
Assigns the measurement to trace 2.
Manual operation:
See
[SENSe:]POWer:ACHannel:PRESet
<Setting>
This command adjusts the frequency span, the measurement bandwidths and the
detector as required for the number of channels, the channel bandwidths and the chan-
nel spacings selected in the active power measurement. If necessary, adjacent-chan-
nel power measurement is switched on prior to the adjustment.
To obtain correct results, a complete sweep with synchronization to the end of the
sweep must be performed after the adjustment. Synchronization is possible only in the
single sweep mode.
The result is queried with the
CALCulate<n>:MARKer<m>:FUNCtion:POWer:
command.
Parameters:
<Setting>
ACPower | CPOWer | MCACpower | OBANdwidth | OBWidth |
CN | CN0
Example:
POW:ACH:PRES ACP
Sets the frequency span, the measurement bandwidths and the
detector as required for the ACLR measurement.
INIT:CONT OFF
Switches over to single sweep mode.
INIT;*WAI
Starts a sweep and waits for the end of the sweep.
CALC:MARK:FUNC:POW:RES? ACP
Queries the result of the adjacent-channel power measurement.
Manual operation:
See
See
Remote Commands in Spectrum Analyzer Mode