Remote Control
R&S
®
ESRP
685
User Manual 1175.7445.02 ─ 08
Example:
CALC:LIM3:UPP:SHIF 20
Shifts all Y values of limit line 3 by 20 limit line units, e.g. dB.
Usage:
Event
CALCulate<n>:LIMit<k>:UPPer:SPACing
<InterpolType>
This command selects linear or logarithmic interpolation for the upper limit line.
Suffix:
<n>
.
irrelevant
<k>
1...8
Selects the limit line.
Parameters:
<InterpolType>
LINear | LOGarithmic
*RST:
LIN
Example:
CALC:LIM:UPP:SPAC LIN
CALCulate<n>:LIMit<k>:UPPer:STATe
<State>
This command turns a lower limit line on and off.
Before you can use this command, you have to select a limit line with
.
The limit check is turned on separately. For more information see
Suffix:
<n>
.
irrelevant
<k>
1...8
Selects the limit line.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:LIM4:UPP:STAT ON
Switches on limit line 4 (upper limit).
CALCulate<n>:LIMit<k>:UPPer:THReshold
<Threshold>
This command defines a threshold for relative limit lines.
The R&S
ESRP uses the threshold for the limit check, if the limit line violates the
threshold.
Suffix:
<n>
.
irrelevant
Remote Commands in Receiver Mode