Remote Control
R&S
®
ESRP
670
User Manual 1175.7445.02 ─ 08
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:DELT1 ON
Switches marker 1 to delta marker mode.
Manual operation:
See
"Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
See
CALCulate<n>:DELTamarker<m>:TRACe
<TraceNumber>
This command selects the trace a delta marker is positioned on.
The corresponding trace must have a trace mode other than "Blank".
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<TraceNumber>
1 ... 6
Trace number the marker is positioned on.
Example:
CALC:DELT3:TRAC 2
Assigns delta marker 3 to trace 2.
Manual operation:
See
See
See
CALCulate<n>:DELTamarker<m>:X
<Position>
This command positions a delta marker on a particular coordinate on the x-axis.
The position is an absolute value.
You can enter a relative value, if you activate relative entry first with
.
If fixed reference is active, the position is relative to the fixed reference
(
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed[:STATe]
is ON). A query
always returns the absolute position.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<Position>
0 to maximum frequency or sweep time
Example:
CALC:DELT:X?
Outputs the absolute frequency/time of delta marker 1.
Remote Commands in Receiver Mode