Remote Control
R&S
®
ESRP
650
User Manual 1175.7445.02 ─ 08
Example:
CALC:PEAK:METH SUBR
CALC:PEAK:SUBR 25
Selects a subrange search and defines 25 subranges.
Manual operation:
See
See
"No of Subranges / Peaks per Subrange"
CALCulate<n>:PEAKsearch|PSEarch:SUBRanges:PCOunt
<Peaks>
This command defines the number of peaks to be found in each subrange.
Before you can use this command, you have to select the subrange peak search mode
with
CALCulate<n>:PEAKsearch|PSEarch:METHod
.
Parameters:
<Peaks>
Number of peaks in one subrange.
Note that the maximum number of peaks is 500. Thus, the maxi-
mum number of peaks per subrange depends on the number of
subranges you have defined.
Range:
1 to depends on the number of subranges
*RST:
1
Example:
CALC:PSE:METH SUBR
CALC:PSE:SUBR 20
CALC:PSE:SUBR:PCO 5
Looks for 5 peaks in each of 20 subranges.
Manual operation:
See
"No of Subranges / Peaks per Subrange"
11.4.6.4
Peak Lists
Commands useful to read peak lists described elsewhere:
●
DISPlay[:WINDow<n>]:TRACe<t>:SYMBol
.......................................................................650
............................................................................................ 651
DISPlay[:WINDow<n>]:TRACe<t>:SYMBol
<Symbol>
This command turns the peak labels in the diagram on and off.
Parameters:
<Symbol>
CROSs
Each peak is labelled by a symbol. The symbol and its color
depend on the trace the peak is on.
OFF
Peak labels are off.
*RST:
OFF
Example:
DISP:TRAC:SYMB CROS
Remote Commands in Receiver Mode