Remote Control
R&S
®
ESRP
805
User Manual 1175.7445.02 ─ 08
Parameters:
<Delay>
Range:
zero span: -sweeptime (see data sheet) to 30 s;
span: 0 to 30 s
*RST:
0 s
Example:
TRIG:HOLD 500us
Manual operation:
See
TRIGger<n>[:SEQuence]:IFPower:HOLDoff
<Value>
This command sets the holding time before the next IF power trigger event.
Note
: Using gating with gate mode "level" (see
on page 865) and an IFP trigger (see
TRIGger<n>[:SEQuence]:LEVel:IFPower
on page 806), the holdoff time for the IFP trigger is ignored for frequency sweep, FFT
sweep, zero span and IQ mode measurements.
Suffix:
<n>
.
irrelevant
Parameters:
<Value>
*RST:
150 ns
Example:
TRIG:SOUR IFP
Sets the IF power trigger source.
TRIG:IFP:HOLD 200 ns
Sets the holding time to 200 ns.
Manual operation:
See
TRIGger<n>[:SEQuence]:IFPower:HYSTeresis
<Value>
This command sets the limit that the hysteresis value for the IF power trigger has to fall
below in order to trigger the next measurement.
Suffix:
<n>
.
irrelevant
Parameters:
<Value>
*RST:
3 dB
Example:
TRIG:SOUR IFP
Sets the IF power trigger source.
TRIG:IFP:HYST 10DB
Sets the hysteresis limit value.
Manual operation:
See
TRIGger<n>[:SEQuence]:LEVel[:EXTernal]
<TriggerLevel>
This command sets the level of the external trigger source in Volt.
Remote Commands in Spectrum Analyzer Mode